Detail KLS Martin ME MB 3
KLS Martin ME MB 3
ME MB 3 technical data
Power requirements 220– 240 V, 50– 60 Hz; or 100– 127 V, 50– 60 Hz
( adjustable by Technical Service via plug-in jumper inside unit)
Power input without HF output: approx. 26 VA
at max. power output: approx. 800 VA
Protection class I
Classification acc. to MDD II b
Leakage currents LF and HF in compliance with IEC 601, Part 2-2
Type CF; defibrillation-proof
Nominal frequency 450 kHz
Modulation frequency 30 kHz
HF output power:
Type of current Power Crest factor Voltage
Cut 1 ( pure) max. 400 W at 300 © 1.6 at 300 © max. 2, 300 Vpp
Cut 2 ( blend) max. 300 W at 300 © 1.9 at 300 © max. 2, 500 Vpp
Cut 3 ( super blend) max. 200 W at 300 © 2.4 at 300 © max. 2, 700 Vpp
Forced Prep max. 120 W at 200 © 4.9 at 200 © max. 3, 900 Vpp
Endo Mode max. 100 W at 200 © 1.6 at 200 © max. 2, 800 Vpp
Contact coagulation max. 250 W at 200 © 3.4 at 200 © max. 3, 200 Vpp
Spray coagulation max. 120 W at 300 © 5.6 at 300 © max. 9, 000 Vpp
Bipolar coagulation max. 100 W at 100 © 2.1 at 100 © max. 600 Vpp
Duty type intermittent 10 s / 30 s = duty factor of 25%
Line fuses 220– 240 V: T 4 A ( slow-blow)
100– 127 V: T 8 A ( slow-blow)
Sound level HF current activation: 40 - 65 dB ( adjustable)
Alarm: 65 dB( A)
Weight 8.6 kg
Interference suppression Limits in conformity with EN 55011
Interference immunity in conformity with IEC 801
Dimensions Width: 405 mm; height: 135 mm; depth: 380 mm
CE-marking According to 93/ 42/ EEC